ISO 10993 is a series of biocompatibility tests; for a silicone part, the tests depend on contact type and duration. A surface device with limited skin contact typically needs cytotoxicity (10993-5), sensitization (10993-10) and irritation (10993-10/23); a long-term implant needs far more. Start by defining the contact category, then select the tests.
Contact category decides the tests
| Contact type | Typical tests |
|---|---|
| Surface, skin (< 24 h) | Cytotoxicity, sensitization, irritation |
| Mucosal / breached surface | + implantation or systemic tests |
| Blood / tissue contact | + hemocompatibility, implantation |
| Implant, long term | Full battery per ISO 10993-1 |
What a supplier should provide
A silicone supplier does not "certify" your device — you or your lab run the ISO 10993 battery on the finished part. What the supplier can provide is a material-grade LSR with an existing ISO 10993 test summary (cytotoxicity, sensitization, irritation), plus USP Class VI and documentation that lets you leverage the material data. That is the difference between buying silicone and buying a documented medical material.