±0.02 mm
Critical-dimension tolerance held on the finished diaphragm sealing lip.
A diaphragm is the moving sealing element in pumps, valves, regulators and medical fluid paths — a thin, flexible silicone membrane that flexes to actuate or seal. It is distinct from a static gasket (compressed between two non-moving faces) and a round O-ring. Because a diaphragm deforms on every cycle, its sealing lip has to stay dimensionally stable across thousands of flex cycles — and that is exactly the control we document and verify.
A thin, flexible silicone membrane that flexes to actuate, seal or regulate fluid and gas.
Platinum-cured LSR (USP Class VI / ISO 10993), food-grade or industrial silicone.
Flat, domed, convoluted or rolling — molded to your drawing with tight lip tolerances.
Send a drawing or sample — we review material, thickness and tolerance before quoting.
Free review · no obligation
A diaphragm must flex repeatedly while keeping a leak-tight barrier against liquids and gases. The failure modes are different from a static seal: a lip that drifts by even a few hundredths of a millimetre causes the device to leak or bind. We design and mold diaphragms with the wall thickness, gate placement and sealing-lip geometry that stay stable over the part's working life.
Diaphragm material is chosen for flex-life, chemical contact and temperature, not just sealing. The grade below depends on what the membrane touches and how many cycles it must survive.
| Diaphragm application | Recommended grade | Typical hardness |
|---|---|---|
| Medical fluid-path / respiratory | Medical-grade platinum-cured LSR (USP Class VI / ISO 10993) | 40 – 50A |
| Food & beverage / water | Food-grade silicone (FDA / LFGB) | 50 – 60A |
| Industrial pump / regulator | Solid silicone rubber (RoHS / REACH) | 60 – 70A |
| Fuel & solvent contact | Fluorosilicone (FVMQ) | 60 – 75A |
Compression-molded flat and domed membranes, and injection-molded LSR diaphragms, with sealing-lip thickness tuned to your stroke and pressure.
Critical lip geometry is held with mold cavities machined to ±0.008 mm (CNC) and ±0.003 mm on EDM fine features, so the tool is not the limiting factor.
Medical diaphragms are molded and packed in an ISO 14644-1:2015 Class 7 cleanroom with robotic handling and full lot traceability.
We developed a medical liquid-silicone diaphragm seal end to end — drawing review, DFM, cold-runner tooling, mold trials and dimension verification — with a measured process capability of Cpk 1.24 on the critical inner-diameter dimension (study LSR-MED-007).
Critical-dimension tolerance held on the finished diaphragm sealing lip.
Verified on the critical inner diameter, reported honestly against the 1.33 target.
Molded and packed in an ISO Class 7 cleanroom with robotic handling.
Platinum-cured LSR on S136 tool steel with a cold-runner feed system.
Read the full diaphragm case study — from drawing to mass production →
Send your drawing, sample or dimensions. We will review material, thickness and tolerance before quoting.
Compare all sealing solutions →
Request a Quote Industrial RFQ →