If your silicone part touches a drug product, IV fluid or the human body, regulators will ask about extractables and leachables (E&L). This article explains what the terms mean, which standards apply, and what documentation you should ask your silicone supplier for — in plain engineering language.

Extractables vs. Leachables: The Key Distinction

The two terms are often confused, but they describe different stages of the same investigation:

TermWhat it measuresWhen it is run
ExtractablesChemical substances pulled out of the material under aggressive lab conditions (strong solvents, heat, extended time)Early, as a "worst-case" screening of what the material could release
LeachablesSubstances that actually migrate into the real drug product or fluid under normal-use conditionsLater, on the finished device or drug, as part of stability studies

Think of extractables as the full list of "what might come out," and leachables as the smaller, real-world list of "what actually does." A material with a broad extractables profile can still have an acceptably small leachables profile once real contact conditions are applied.

The Standards That Govern E&L

  • ISO 10993-18 — the core standard for chemical characterization of medical-device materials. It sets the framework for designing an extractables study: solvent selection, extraction conditions and analytical methods.
  • USP <1663> and <1664> — pharmaceutical-focused chapters on assessing extractables and leachables for packaging and delivery systems (US Pharmacopeia).
  • ISO 10993-1 — the parent standard that drives when chemical characterization is required as part of a biological-evaluation plan.
  • ICH Q3C / Q3D — residual-solvent and elemental-impurity limits that feed into E&L risk assessment.

For silicone specifically, the relevant concerns are typically residual siloxanes (low-molecular-weight cyclic or linear oligomers such as D4, D5, D6), platinum-catalyst residues, peroxide by-products, and any compounding additives. Platinum-cured LSR generally produces a cleaner extractables profile than peroxide-cured solid silicone, which is one reason it is preferred for many drug-contact and long-term implantable-adjacent applications.

What a Solvent Panel Typically Includes

Because no single solvent mimics every fluid, extractables studies run a panel that brackets a range of polarities:

  • Polar / aqueous — water, buffered saline, or an alcohol–water mix.
  • Semi-polar — isopropanol, or ethanol/water mixtures.
  • Non-polar — hexane or another hydrocarbon, used as a worst-case for lipophilic extraction.

The choice of solvents and extraction conditions (temperature, time, surface-area-to-volume ratio) is dictated by the intended clinical use — this is where a well-designed study avoids both false negatives and unnecessary over-testing.

Analytical Techniques Used

  • GC-MS — identifies and quantifies volatile and semi-volatile organics, including cyclic siloxanes.
  • LC-MS / HPLC — covers non-volatile and thermally labile compounds.
  • ICP-MS / ICP-OES — measures elemental impurities and metal residues.
  • FTIR and gravimetric NVR — non-volatile-residue screening and material fingerprinting.

See our silicone testing methods guide for a deeper look at ICP-OES and GC-MS specifically.

What to Request from Your Silicone Supplier

Before you commit, ask for documentation that shows the material has been characterized, not just "certified":

  • A material extractables report with the solvent panel, conditions and detected compounds listed — not just a pass/fail summary.
  • ISO 10993-18 chemical-characterization data, or a statement of readiness to support one.
  • Biocompatibility test data to USP Class VI and/or the relevant ISO 10993 endpoints for your contact duration.
  • The cure system used (platinum vs. peroxide) — this materially affects the extractables profile.
  • Raw-material traceability to a named silicone feedstock (Elkem, Dow, DuPont or equivalent) so results are reproducible lot-to-lot.

How CYES SILICONE Supports E&L

We compound and mold medical-grade and platinum-cured LSR in-house, source silicone feedstock from suppliers including Elkem, Dow and DuPont, and test every incoming raw-material lot before production. For drug-contact and medical projects we support chemical-characterization programs by providing material identity, cure-system detail and traceable documentation to your ISO 10993-18 or USP <1663> study — so your regulatory file is built on a characterized material rather than an assumption. See our quality documents page for the full documentation set we can provide.